VASC-ALERT ACCESS SURVEILLANCE CASE STUDY 6
This patient had multiple high readings and alerts issued by Vasc-Alert. In the above VAPR graphs each vertical green line indicates the average VAPR for the dialysis session. The horizontal blue line is a pre-set threshold value. The red line is a moving average which is used to visualize the trend of the graph more easily. The red arrow indicates when an intervention took place. The dates on the graph are in month/day format. There were no clinical signs or symptoms that were indicative of access dysfunction or stenosis (e.g. no increase in venous pressure, no excessive bleeding, no decrease in blood flow, no difficulty in cannulation, no decrease in Kt/V), but the patient had consistent alerts and an increasing trend of VAPR results above threshold. The patient was sent for a fistulogram on 8/23/06 based solely on Vasc-Alert results and was determined to have significant stenosis, so an angioplasty was performed. The red arrow on the graph indicates the drop in VAPR values below the threshold immediately after a fistulogram and angioplasty. PROCEDUREThe patient's left upper arm dialysis shunt is accessed near its presumed origin near the antecubital region. Through this, contrast is injected and images of the shunt are taken as well as the major draining veins of the left upper extremity. Using the same approach, imaging is done of the major draining veins of the left upper extremity. Using the same approach, angioplasty is done of an outflow lesion described below. FINDINGS
The access shunt is patent with no stenosis but there is moderate deterioration of its mid portion with small umbilications at previous access puncture sites. There is tight narrowing at the venous outflow which is to the axillary vein with 80% concentric narrowing here.
The patient was sent for access study based on Vasc-Alert results and found to have a hemodynamically significant stenosis of 80% that was treated by angioplasty. Shortly after intervention the patient's VAPR results returned to normal and fell below the threshold. |
| © 2008, 2009, 2010, 2011, 2012 Vasc-Alert. All rights reserved. | About Us | Privacy Statement | |